AAWSAP DIRD Biosensors and BioMEMS A Survey of the Present Field March 31 2010
UNCLASSIFIED/ /P'OR: OP'P'l@IAL WS& 8NI.¥
Thin Mylar Substrate
Figure 15. Thermopile Glucose Sensor (left) With Functional Illustration (right) are Made Using Photolithographic Techniques. Th is larger device is about 8 mm in lengt h but has approxi mat el y 50 m icron minim um geometries. (Towe et al. )
At ASU the glucose sensor has been constructed on a thin mylar plastic substrate and so can be curved into a cyl inder to form a small tube. This tube is then sensitive to glucose concentrations on its outside exposure to the blood stream. Figure 16 shows a photograph and an illustration of a glucose sensor implemented in the form of a catheter. The electrical output signal is routed by wires down the length of the catheter for a remote readout. The device was implanted in a pig over a short duration and the decrease in blood glucose concentrations in response to admin istration of insulin were recorded and are shown in the right hand panel of Figure 16.
10 min
Figure 16. Photograph and Illustration of the Thermopile Glucose Sensor in a Catheter. Its confi gurat ion is shown ( left) and its response in detecting ch anges in blood glucose in a pig as a result of insul in adm inistration to t he animal. (ASU resea rch.) Towe et al.
Th is glucose sensor approach has been found to have limitations in stability when introduced into tissue or blood. Primarily the problems, as with many sensors, arise from the system chemistry and not as much from the electronic portion of the sensor . The glucose oxidase enzyme slow ly decays over time and thus the sensor sensitivity and calibration drifts making it eventually unusable. These problems have no easy solution. Investigators have been working with various forms of blood chemistry sensors for forty years or more. Microfabricated
19