Figure 13. A MEMS Thermopile Glucose Sensor (Towe et al., Arizona State University)
Heat energy released in the gel is measured by the thermopile and is characterized in terms of kiloJoules (kJ) per mole of glucose consumed. At concentrations of glucose that are common in the blood stream, the temperature r ise is very small, on t he order of a hundredth of a degree (10 millidegrees) . Measuring this temperature change in the presence of ambient body heat or room temperature changes of a few degrees is a daunting task, but can be done. The measurement of the small reaction temperature in the face of possible ambient temperature shifts is accomplished by the use of a differential temperature measurement system. This sensor employs what is known as the Seebeck Effect whereby dissimilar metals heated at their junction produce a voltage that is proportional to temperature. This concept is shown in Figure 14. This principle is then implemented as an array of metal junctions that when connected in series produce a higher output in order to detect smaller temperature changes.