AAWSAP DIRD Biosensors and BioMEMS A Survey of the Present Field March 31 2010

UNCLASSIFIED/ /P'OR: OP'Pl@IAL WS& 0NI.¥

An enzyme-gel is applied in a thin layer over a electrically biased metal electrode. An electrical current flows through the electrode when the hydrogen peroxide produced by the glucose reaction is decomposed and causes an electrical current to flow. If there is a fixed concentration of glucose present, a corresponding amount of peroxide will be produced, and this can be measured by the electrical current flow. The amount of peroxide measured is an indicator of the glucose concentration. The actual amount of analyte consumed by a small BioMEMS sensor is exceedingly small and has no significant effect on the local concentration.
Figure 11 shows a MEMS implementation of a blood glucose sensor using this kind of approach (Advanced Biosensors Inc). The fine needle geometry is generally meant for insertion into tissue. This device is comprised of multiple independent small sensors and is actually a system of components that adds signal processing and interface electronics to allow its communication with insulin pumps for automated delivery of insulin based on patient need.

Advanced BioSensors blood glucose sensor
- Flip chip assembly
AS ICs , pow er sources and circui componen s
Separa e sensors , moun w i h electrical connec ions , add he biocompa ible polym er, assemble he sensor pa ch.
The sensor based on glucose oxidase reacti on wi h blood plasma from ca pill aries in he dermi s.
- biochemical coa ing - sensor in place 3 - 7 days
- Docking par for elec ronics assembly w i h amplifi er, ADC , w ireless transceiver , pow er source
- encryp ed digi al da a to w ri stwa ch-sized morn or/ record er module
In fu ure combine the CG S wi h an insulin pump o produce a closed-loop sys em-an artifici al pa n creas .

' . • -

•

..•· -~-s

-

•

-

-

- 2-

long x 200 u Ide

Figure 11. An Experimental MEMS Blood Glucose Sensor
Glucose sensors based on electrical measurement of hydrogen peroxide products have been found to have the significant problem of a drifting baseline over time. It is difficult to maintain calibration of the sensors, and the electrode is affected by proteins and other substances in the blood stream. Although this sensing approach has had some success, its longevity in the body is much less than desired; sensors based on this approach decay alter a few days of use.
THERMOPILE IMPLANTABLE GLUCOSE SENSORS
Towe et al. at Arizona State University have been working on an improved approach to blood glucose sensing. It depends on the fact that the glucose oxidase reaction generates a small amount of heat as indicated in the above chemical reaction. The