AAWSAP DIRD Biosensors and BioMEMS A Survey of the Present Field March 31 2010

UNCLASSIFIED/ /P'OR: OP'P'l@IAL WS& 0Nk¥
drift and offset in the reading and so shifting the calibration. Depending on its exact construction some versions however can be effective for as much as a few days. Figure 8 shows the latest in microdevice oxygen sensors. (This particular version is not specifically a MEMS device since it is not made by photolithography and it is neither .§lectrical nor mechanical in nature.) It represents however the trend in oxygen sensors towards optical measurement techniques that are evolving towards optical systems based on MEMS. These sensors can be implemented in small sizes with optical fibers that transmit and receive light from the sensor surface at the tip. Variations on this basic principle are used for indwelling blood oxygen catheters during surgery.
fiber op cs silicone cover
Figure 8. A Fiber Optical Oxygen Sensor Showing Two Optical Fibers Mated to Plastic Prisms That are Coated With an Oxygen-Sensor Coating, Usually a Fluorescent Ruthenium Compound ( http://www.imtek.de/content/projekte_en. php?ls= 11)
The oxygen sensors respond to changes in absorption of oxygen into the special coating at the t ip, or in some versions, they respond to fluorescence of a thin oxygen-sensitive coating placed on the prism surface. The secret to good performance lies in the proprietary selection of the specific chemical species selected for the coating. Blue light is usually transmitted down the optical fiber and the coating glows orange in proportion to the amount of oxygen present. The light from the glow is conducted back toward the supporting instrumentation system. The intensity of the orange light then indicates oxygen levels in the solution. The particular sensor shown in Figure 8 uses two optical fibers. The one on the left side is not sensitive to oxygen, but rather acts as a reference system that is used to compensate for changes temperature, ambient light, and other system variables.
12 UNCLASSIFIED/ {FOR OFFICIO! !PEii QDlk¥